FDA Device recall Z-0686-2015
PKG, GRASPER FORCEPS, ATRAUMATIC, P/N 0250080323
- FDA Device
- Class II
- terminated
- Published 2014-12-24
On 2014-12-24 the FDA classified a Class II recall of PKG, GRASPER FORCEPS, ATRAUMATIC, P/N 0250080323 by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0686-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-17 |
| Published | 2014-12-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Endoscopy |
| Distribution | US |
Product
PKG, GRASPER FORCEPS, ATRAUMATIC, P/N 0250080323. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Identifiers
| code_info | 747135 840073 840364 846994 1141383 1141658 1142912 1143122 1143498 1144242 1144502 1240415 1241513 124617…747135 840073 840364 846994 1141383 1141658 1142912 1143122 1143498 1144242 1144502 1240415 1241513 1246170 1340543 1341580 1342751 1346234 642223 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0686-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.