RecallRadar

FDA Device recall Z-0685-2017

Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with…

On 2016-12-07 the FDA classified a Class III recall of Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with… by Microport Orthopedics, citing revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to pr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0685-2017
ClassificationClass III
Statusterminated
Initiated2016-06-21
Published2016-12-07
Last updated2026-09-13 11:54 UTC
CompanyMicroport Orthopedics
DistributionUS

Product

Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head

Reason

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Identifiers

code_info
REF PHA04702 Serial Numbers: PHA04702, 087461467, 089909592, 106368684, 106379196, 106381827, 109949208, 129989153, 02…REF PHA04702 Serial Numbers: PHA04702, 087461467, 089909592, 106368684, 106379196, 106381827, 109949208, 129989153, 0201018636, 0301049040 REF PHA04704 Serial Numbers: 010989155, 027373842, 038461468, 057433172, 077453350, 089924904, 106379197, 106380366, 106381751, 0201018637 REF PHA04706 Serial Numbers: 1414640, 010990534, 017400915, 037415861, 039755749, 047420835, 047424536, 077459275, 087461469, 089906466, 0101034747, 106379879, 106381773, 0301084459, 0401100079, 0501122040, 0801189074 REF PHA04708 Serial Numbers: 010990535, 017400568, 037420836, 047424415, 077455537, 087461470, 089901226, 106380367, 106381752, 0311304891, 0401118999, 0601163086, 0711360334 REF PHA04710 Serial Numbers: 1372615, 1413361, 017400673, 037415862, 037420837, 047423789, 047424535, 077457450, 077460608, 089901146, 098695744, 098695746, 106380368, 106381542, 106381753, 108695745, 109944906, 116380977, 119983782, 129990536, 0201016259, 0301076759, 0311303536, 0311305118, 0401089452

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0685-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.