RecallRadar

FDA Device recall Z-0685-2013

DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expediu…

On 2013-01-23 the FDA classified a Class II recall of DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expediu… by Depuy Spine, citing A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specifica….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0685-2013
ClassificationClass II
Statusterminated
Initiated2012-07-24
Published2013-01-23
Last updated2026-09-13 11:53 UTC
CompanyDepuy Spine
DistributionUS

Product

DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.

Reason

A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.

Identifiers

code_infoAll lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0685-2013%22