RecallRadar

FDA Device recall Z-0681-2015

PKG, GRASPER FORCEPS, SERRSTED FENESTRATED, P/N 0250080318

On 2014-12-24 the FDA classified a Class II recall of PKG, GRASPER FORCEPS, SERRSTED FENESTRATED, P/N 0250080318 by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0681-2015
ClassificationClass II
Statusterminated
Initiated2014-11-17
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

PKG, GRASPER FORCEPS, SERRSTED FENESTRATED, P/N 0250080318. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Identifiers

code_info
543367 544893 842088 845454 941849 942135 942844 943742 945442 1040419 1041098 1042499 1042929 1043777 10…543367 544893 842088 845454 941849 942135 942844 943742 945442 1040419 1041098 1042499 1042929 1043777 1044526 1044527 1045033 1045372 1047071 1141219 1141973 1142134 1142913 1143973 1144935 1145348 1145796 1146957 1147046 1241906 1243270 1244187 1246179 1341468 1341557 1343701 1344283 1344526 1345529 1346236 1346680 1440403 1440910 1442176 1442809 1443395 1443652 646462

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0681-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.