FDA Device recall Z-0678-2018
Nutrisafe 2 35 mL Syringe for Safety Connection
- FDA Device
- Class II
- terminated
- Published 2018-02-28
On 2018-02-28 the FDA classified a Class II recall of Nutrisafe 2 35 mL Syringe for Safety Connection by Vygon Mfg Dba Churchill Medical Systems, citing vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product,….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0678-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-07-24 |
| Published | 2018-02-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Vygon Mfg Dba Churchill Medical Systems |
| Distribution | n/a |
Product
Nutrisafe 2 35 mL Syringe for Safety Connection. Product Code: 1015.352M
Reason
Vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the Tyvek may not be coated completely causing gaps in the adhesive used to seal the product packaging. This potential lack of adhesive may affect the sterility of the product.
Identifiers
| code_info | Lot Number: 1705060D |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0678-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.