FDA Device recall Z-0670-2022
Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301
- FDA Device
- Class II
- terminated
- Published 2022-03-02
On 2022-03-02 the FDA classified a Class II recall of Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301 by Biomet, citing the products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion str….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0670-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-01-07 |
| Published | 2022-03-02 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Biomet |
| Distribution | US |
Product
Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301
Reason
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Identifiers
| code_info | Lot Number: 853650, 853810 UDI: (01)00880304468030(17) 290122(10)853650; (01)00880304468030(17) 290211(10)853810 |
|---|