FDA Device recall Z-0670-2017
sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique De…
- FDA Device
- Class II
- terminated
- Published 2016-12-07
On 2016-12-07 the FDA classified a Class II recall of sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique De… by Baxter Healthcare, citing A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblie….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0670-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-09-28 |
| Published | 2016-12-07 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Baxter Healthcare |
| Distribution | US |
Product
sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.
Reason
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Identifiers
| code_info | Contact CDRH for list of affected serial numbers (5976 units) |
|---|