RecallRadar

FDA Device recall Z-0669-2018

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small…

On 2018-02-28 the FDA classified a Class II recall of LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small… by Synthes, citing may have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is d….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0669-2018
ClassificationClass II
Statusterminated
Initiated2017-08-07
Published2018-02-28
Last updated2026-09-13 11:54 UTC
CompanySynthes
Distributionn/a

Product

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments

Reason

May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.

Identifiers

code_infoLot numbers FT00227 9972971 H028400 H161606 H288454

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.