FDA Device recall Z-0669-2018
LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small…
- FDA Device
- Class II
- terminated
- Published 2018-02-28
On 2018-02-28 the FDA classified a Class II recall of LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small… by Synthes, citing may have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is d….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0669-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-08-07 |
| Published | 2018-02-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Synthes |
| Distribution | n/a |
Product
LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments
Reason
May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.
Identifiers
| code_info | Lot numbers FT00227 9972971 H028400 H161606 H288454 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0669-2018%22