FDA Device recall Z-0667-2013

Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns&…

On 2013-01-23 the FDA classified a Class II recall of Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns&… by Terumo Cardiovascular Systems, citing flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specific….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0667-2013
ClassificationClass II
Statusterminated
Initiated2013-01-07
Published2013-01-23
Last updated2026-09-13 11:53 UTC
CompanyTerumo Cardiovascular Systems
DistributionUS

Product

Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery

Reason

Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.

Identifiers

code_infoLot number: 0677284 and Catalog number: 13020

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.