FDA Device recall Z-0666-2017
Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator support to adults and pedia…
- FDA Device
- Class II
- terminated
- Published 2016-12-07
On 2016-12-07 the FDA classified a Class II recall of Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator support to adults and pedia… by Hamilton Medical, citing oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0666-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-09-26 |
| Published | 2016-12-07 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Hamilton Medical |
| Distribution | US |
Product
Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator support to adults and pediatrics, and optionally infants and neonates. Intended areas of use include: MRI Dept., Intensive, intermediate, emergency and long-term acute care as well as transfer of patients within a hospital.
Reason
Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.
Identifiers
| code_info | Serial Numbers between 2001 and 2103 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.