RecallRadar

FDA Device recall Z-0664-2013

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for…

On 2013-01-16 the FDA classified a Class II recall of Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for… by Siemens Healthcare Diagnostics, citing firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an applic….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0664-2013
ClassificationClass II
Statusterminated
Initiated2012-12-10
Published2013-01-16
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionCA, FL, IL, KS, MA, MI, MN, MS, NC, NM, NY, OH, OR, PA, TN, TX, UT, WA

Product

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for use in the quantitative analysis of Phenobarbital in human serum or plasma.

Reason

Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an application if the AU(R) System is used, the issue may be experienced. Firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. This is an expansion of

Identifiers

code_infoEmit(R) 2000 Phenobarbital Assay (OSR4D229) lots 1237 and 1264 and/or Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots E1, E2, and E3.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0664-2013%22