RecallRadar

FDA Device recall Z-0663-2022

Philips Respironics V60 Plus Ventilator Part Number 1138747

On 2022-03-09 the FDA classified a Class I recall of Philips Respironics V60 Plus Ventilator Part Number 1138747 by Respironics California, citing A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0663-2022
ClassificationClass I
Statusongoing
Initiated2022-01-24
Published2022-03-09
Last updated2026-09-13 11:56 UTC
CompanyRespironics California
DistributionAR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN, TX

Product

Philips Respironics V60 Plus Ventilator Part Number 1138747

Reason

A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).

Identifiers

code_info
The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part…The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1138747 Serial Numbers/UDI Code: MX0016240/ (01)00884838090996(21)MX0016240 MX0016241/ (01)00884838090996(21)MX0016241 MX0016242/ (01)00884838090996(21)MX0016242 MX0016243/ (01)00884838090996(21)MX0016243 MX0016244/ (01)00884838090996(21)MX0016244 MX0016245/ (01)00884838090996(21)MX0016245 MX0016246/ (01)00884838090996(21)MX0016246 MX0016247/ (01)00884838090996(21)MX0016247 MX0016248/ (01)00884838090996(21)MX0016248 MX0016522/ (01)00884838090996(21)MX0016522 MX0016523/ (01)00884838090996(21)MX0016523 MX0016525/ (01)00884838090996(21)MX0016525 MX0016527/ (01)00884838090996(21)MX0016527 MX0016529/ (01)00884838090996(21)MX0016529 MX0016531/ (01)00884838090996(21)MX0016531 MX0016532/ (01)00884838090996(21)MX0016532 MX0016533/ (01)00884838090996(21)MX0016533 MX0016534/ (01)00884838090996(21)MX0016534 MX0016535/ (01)00884838090996(21)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0663-2022%22