FDA Device recall Z-0660-2022
The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless…
- FDA Device
- Class II
- ongoing
- Published 2022-03-02
On 2022-03-02 the FDA classified a Class II recall of The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless… by Balt Usa, citing due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mi….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0660-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-01-04 |
| Published | 2022-03-02 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Balt Usa |
| Distribution | US |
Product
The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil delivery pusher with a radiopaque distal positioning marker and proximal fluorosafe marker. The Optima Coil delivery pusher is 185cm in length. it is designed for use with the XCEL Detachment Controller. The XCEL Detachment Controller comes pre-loaded with batteries and is a sterile, handheld, single-patient-use device which should not to be reused, re-sterilized, opened or tampered with. The Optima Coil System and XCEL Detachment Controller(s) are sold separately. The Optima Coil achieves detachment by an internal heater element, which is powered by the XCEL Detachment Controller.
Reason
Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.
Identifiers
| code_info | Model/REF: OPTI0204CSF10 Lot Number: F210500134 UDI Code: 00818053026010 Model/REF: OPTI0407CSF10 Lot Number: F210500135 UDI Code: 00818053026096 |
|---|