FDA Device recall Z-0660-2018

Aspire HD Plus, Aspire HD-s (for FDR-2000AWS) The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire ND-s (FDR M…

On 2018-02-28 the FDA classified a Class II recall of Aspire HD Plus, Aspire HD-s (for FDR-2000AWS) The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire ND-s (FDR M… by Fujifilm Medical Systems U, citing FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and C….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0660-2018
ClassificationClass II
Statusterminated
Initiated2017-08-04
Published2018-02-28
Last updated2026-09-13 11:54 UTC
CompanyFujifilm Medical Systems U
Distributionn/a

Product

Aspire HD Plus, Aspire HD-s (for FDR-2000AWS) The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire ND-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.

Reason

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Identifiers

code_info2000AWS. Released Software Versions: V5.1, V6.0, V6.1. Not Released Software Versions: V5.0, V5.2, V7.0.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.