FDA Device recall Z-0660-2013
PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- FDA Device
- Class II
- terminated
- Published 2013-01-16
On 2013-01-16 the FDA classified a Class II recall of PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented by Biomet, citing biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized con….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0660-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2006-12-12 |
| Published | 2013-01-16 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Biomet |
| Distribution | US |
Product
PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason
Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese
Identifiers
| code_info | Part 113950 Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,28…Part 113950 Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,282590,282600,282620,282630, 282680,282700,282710,489160,489180,489200,489210,489770,489780, 489790,714480,714550,714570,714700,714730,722210,722250,722280, 984770,984780,984790,984800,984810,984820,984830,984840, and 984850 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0660-2013%22