FDA Device recall Z-0659-2018
Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-l000), generates full-field digital mammography imag…
- FDA Device
- Class II
- terminated
- Published 2018-02-28
On 2018-02-28 the FDA classified a Class II recall of Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-l000), generates full-field digital mammography imag… by Fujifilm Medical Systems U, citing FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and C….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0659-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-08-04 |
| Published | 2018-02-28 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Fujifilm Medical Systems U |
| Distribution | n/a |
Product
Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-l000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
Reason
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
Identifiers
| code_info | 1000AWS. Released Software Versions: V5.0, V5.1. Not Released Software Versions: V5.2, V6.0, V6.1, V7.0. |
|---|