RecallRadar

FDA Device recall Z-0658-2022

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

On 2022-03-02 the FDA classified a Class II recall of NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R by Nobel Biocare Usa, citing due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0658-2022
ClassificationClass II
Statusterminated
Initiated2021-12-17
Published2022-03-02
Last updated2026-09-13 11:56 UTC
CompanyNobel Biocare Usa
DistributionUS

Product

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reason

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Identifiers

code_info
Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)2…Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497