RecallRadar

FDA Device recall Z-0655-2024

Abiomed Kit, 23Fr Introducer, 11cm, Sterile Part of Pump Set (1000323)-Introducer(s) are intended for introduction of the Impella Catheter…

On 2024-01-17 the FDA classified a Class II recall of Abiomed Kit, 23Fr Introducer, 11cm, Sterile Part of Pump Set (1000323)-Introducer(s) are intended for introduction of the Impella Catheter… by Abiomed, citing hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to t….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0655-2024
ClassificationClass II
Statusongoing
Initiated2023-11-07
Published2024-01-17
Last updated2026-09-13 11:57 UTC
CompanyAbiomed
DistributionUS

Product

Abiomed Kit, 23Fr Introducer, 11cm, Sterile Part of Pump Set (1000323)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 2000342

Reason

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

Identifiers

code_infoUDI-DI: N/A

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0655-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.