FDA Device recall Z-0652-2017

Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium…

On 2016-11-30 the FDA classified a Class II recall of Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium… by Siemens Healthcare Diagnostics, citing siemens determined that discrepant low calcium recovery of QC and patient samples can occur with Dimension Vista¿ CHEM 1 CAL lots 5GM081 and 5GM082.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0652-2017
ClassificationClass II
Statusterminated
Initiated2016-09-15
Published2016-11-30
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA/CRE2), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System

Reason

Siemens determined that discrepant low calcium recovery of QC and patient samples can occur with Dimension Vista¿ CHEM 1 CAL lots 5GM081 and 5GM082.

Identifiers

code_infoDevice Listing Number: D011448 (Kc110) and D244577 (KC110B) Lot numbers: MGM081 and MGM082