LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended f…

On 2016-11-30 the FDA classified a Class II recall of LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended f… by Diasorin, citing diaSorin is issuing an urgent medical device correction for the LIAISON¿ Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON¿ Estradiol….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0645-2017
ClassificationClass II
Statusterminated
Initiated2016-07-13
Published2016-11-30
Last updated2026-09-13 11:54 UTC
CompanyDiasorin
DistributionUS

Product

LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and laboratory data to assist in the diagnosis and management of conditions involving an excess or deficiency of estradiol. The test has to be performed on the LIAISON Analyzer Family.

Reason

DiaSorin is issuing an urgent medical device correction for the LIAISON¿ Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON¿ Estradiol II Gena assay should not be used to test patients being treated with fulvestrant (Faslodex).

Identifiers

code_infoCatalog Number 31 0680, 31 0680-CN

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.