RecallRadar

FDA Device recall Z-0640-2026

Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, O…

On 2025-12-03 the FDA classified a Class II recall of Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, O… by B Braun Medical, citing potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0640-2026
ClassificationClass II
Statusongoing
Initiated2025-10-29
Published2025-12-03
Last updated2026-09-13 11:57 UTC
CompanyB Braun Medical
Distributionn/a

Product

Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.

Reason

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Identifiers

code_info
1. Catalog Number: 362032; Primary UDI-DI: 04046964182556; Unit of Dose UDI-DI: 04046964182549. Shelf Life: Earliest Ex…1. Catalog Number: 362032; Primary UDI-DI: 04046964182556; Unit of Dose UDI-DI: 04046964182549. Shelf Life: Earliest Exp of Comp or 60mths. 2. Catalog Number: 362033; Primary UDI-DI: 04046964837821; Unit of Dose UDI-DI: 04046964837814. Shelf Life: Earliest Exp of Comp or 36mths. 3. Catalog Number: 362034; Primary UDI-DI: 04046964182570; Unit of Dose UDI-DI: 04046964182563. Shelf Life: Earliest Exp of Comp or 48mths. 4. Catalog Number: 362043; Primary UDI-DI: 04046964182631; Unit of Dose UDI-DI: 04046964182624. Shelf Life: Earliest Exp of Comp or 60mths. 5. Catalog Number: 362050; Primary UDI-DI: 04046964182655; Unit of Dose UDI-DI: 04046964182648. Shelf Life: Earliest Exp of Comp or 60mths. 6. Catalog Number: 362420; Primary UDI-DI: 04046955592715; Unit of Dose UDI-DI: 04046955592708. Shelf Life: 36mths from Start Date. 7. Catalog Number: 362431; Primary UDI-DI: 04046964294013; Unit of Dose UDI-DI: 04046964294006. Shelf Life: Earliest Exp of Comp or 60mths. 8. Catalog Number: 36

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0640-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.