FDA Device recall Z-0640-2016
VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Softwa…
- FDA Device
- Class II
- terminated
- Published 2016-01-20
On 2016-01-20 the FDA classified a Class II recall of VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Softwa… by Ortho Clinical Diagnostics, citing software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing.….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0640-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-03 |
| Published | 2016-01-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
Reason
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.
Identifiers
| code_info | Software Version 3.2 and below; Serial Numbers J36000012 - 36000879 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0640-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.