FDA Device recall Z-0639-2024
Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System
- FDA Device
- Class II
- ongoing
- Published 2024-01-10
On 2024-01-10 the FDA classified a Class II recall of Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System by Zimmer Biomet Spine, citing spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use o….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0639-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-10-16 |
| Published | 2024-01-10 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Zimmer Biomet Spine |
| Distribution | US |
Product
Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System
Reason
Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Identifiers
| code_info | UDI-DI: 00889024002159, Lot: W834061 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0639-2024%22