OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their access…
- FDA Device
- Class II
- terminated
- Published 2022-02-23
On 2022-02-23 the FDA classified a Class II recall of OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their access… by Olympus, citing the detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0638-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-12-13 |
| Published | 2022-02-23 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Olympus |
| Distribution | US |
Product
OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories
Reason
The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent
Identifiers
| code_info | All serial numbers. UDI: 04953170404047 |
|---|
Status history