FDA Device recall Z-0638-2022

OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their access…

On 2022-02-23 the FDA classified a Class II recall of OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their access… by Olympus, citing the detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0638-2022
ClassificationClass II
Statusterminated
Initiated2021-12-13
Published2022-02-23
Last updated2026-09-13 11:56 UTC
CompanyOlympus
DistributionUS

Product

OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories

Reason

The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent

Identifiers

code_infoAll serial numbers. UDI: 04953170404047