Siemens Chek-Stix Combo Pak Catalog Number: 1364 Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemen…

On 2013-01-16 the FDA classified a Class III recall of Siemens Chek-Stix Combo Pak Catalog Number: 1364 Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemen… by Siemens Healthcare Diagnostics, citing when these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0637-2013
ClassificationClass III
Statusterminated
Initiated2012-11-21
Published2013-01-16
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Siemens Chek-Stix Combo Pak Catalog Number: 1364 Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes

Reason

When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).

Identifiers

code_info
Carton/Bottle Lot Number: C9093061C P9093061C C9094081J P9094081J C0100101D P0100101D C0097091E P0097091E C910…Carton/Bottle Lot Number: C9093061C P9093061C C9094081J P9094081J C0100101D P0100101D C0097091E P0097091E C9101022E P9101022E C0099091D P0099091D C0100101E P0100101E C9101022H P9101022H C9101022J P9101022J C0103042B P0103042B C0103042G P0103042G C0106062D P0106062D C9105052C P9105052C