FDA Device recall Z-0636-2021
Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated…
- FDA Device
- Class II
- terminated
- Published 2020-12-23
On 2020-12-23 the FDA classified a Class II recall of Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated… by Kowa Optimed, citing it has been determined that the software distributed with the retinal camera may delete patient information. Specifically, if multiple IDs are delated at the same time, their is a….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0636-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-11-03 |
| Published | 2020-12-23 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Kowa Optimed |
| Distribution | US |
Product
Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated for acquiring, displaying and storing image data from KOWA s mydriatic and non-mydriatic ophthalmic cameras.
Reason
It has been determined that the software distributed with the retinal camera may delete patient information. Specifically, if multiple IDs are delated at the same time, their is a potential for all patient information to be deleted. This could result in delay of treatment.
Identifiers
| code_info | Serial Numbers: 32168900007, 32168900008, 32168900009, 32168900010, 32168900011, 32168900019, 32168900020, 32168900021…Serial Numbers: 32168900007, 32168900008, 32168900009, 32168900010, 32168900011, 32168900019, 32168900020, 32168900021, 32168900022, 32168900025, 32168900035, 32168900036, 32168900037, 32168900044, 32168900045, 32169000052, 32169000144, 32169000145, 32169000152, 32169000153, 32169000166, 32169000167, 32169000168 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0636-2021%22