RecallRadar

FDA Device recall Z-0633-2015

PKG, CLAW FORCEPS, DOUBLE ACTION, P/N 0250080245

On 2014-12-24 the FDA classified a Class II recall of PKG, CLAW FORCEPS, DOUBLE ACTION, P/N 0250080245 by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0633-2015
ClassificationClass II
Statusterminated
Initiated2014-11-17
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

PKG, CLAW FORCEPS, DOUBLE ACTION, P/N 0250080245. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Identifiers

code_info
43240 143822 646341 841071 845057 847851 848110 940791 941351 943551 943748 944835 945745 1040099 1040976…43240 143822 646341 841071 845057 847851 848110 940791 941351 943551 943748 944835 945745 1040099 1040976 1041991 1042472 1042687 1042702 1042706 1141976 1142463 1143133 1144329 1144624 1145286 1145818 1146347 1146686 1240540 1241942 1241943 1242475 1243542 1243543 1244257 1341592 1341593 1341597 1345522 1346223 1346539 1440407 1440672 1440673 1440892 1440893

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0633-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.