RecallRadar

FDA Device recall Z-0630-2016

arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424

On 2016-01-27 the FDA classified a Class I recall of arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424 by Arkray Factory Usa, citing one lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for gl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0630-2016
ClassificationClass I
Statusterminated
Initiated2015-12-18
Published2016-01-27
Last updated2026-09-13 11:54 UTC
CompanyArkray Factory Usa
DistributionFL, IL, KY, MI, NC, NY, OH, TN

Product

arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424. For the in vitro quantitative determination of glucose in human plasma from whole blood samples.

Reason

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

Identifiers

code_infoLot Number EA4M78

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0630-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.