FDA Device recall Z-0630-2014
Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx…
- FDA Device
- Class II
- terminated
- Published 2014-01-15
On 2014-01-15 the FDA classified a Class II recall of Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx… by Stryker Instruments Div, citing the Color Cuff¿ Sterile Disposable Tourniquet Cuffs with Quick-Connect Fitting may have an incorrect end-connector. A customer reported having received a luer connector instead of….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0630-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-12-06 |
| Published | 2014-01-15 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Instruments Div |
| Distribution | MI, OH, TX |
Product
Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx only, STERILE
Reason
The Color Cuff¿ Sterile Disposable Tourniquet Cuffs with Quick-Connect Fitting may have an incorrect end-connector. A customer reported having received a luer connector instead of a quick connect fitting. An incorrect connector may lead to an inoperable system due to incompatibility with the Smart Pump.
Identifiers
| code_info | Lot 50731155 Color code Purple |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0630-2014%22