FDA Device recall Z-0629-2018
Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that sup…
- FDA Device
- Class II
- terminated
- Published 2018-02-21
On 2018-02-21 the FDA classified a Class II recall of Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that sup… by Cincinnati Sub Zero Products, citing due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0629-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-11-27 |
| Published | 2018-02-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Cincinnati Sub Zero Products |
| Distribution | US |
Product
Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm or cool the blood that enters and exits a blood heat exchanger for adult, pediatric or infant patients.
Reason
Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.