RecallRadar

FDA Device recall Z-0629-2018

Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that sup…

On 2018-02-21 the FDA classified a Class II recall of Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that sup… by Cincinnati Sub Zero Products, citing due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0629-2018
ClassificationClass II
Statusterminated
Initiated2017-11-27
Published2018-02-21
Last updated2026-09-13 11:54 UTC
CompanyCincinnati Sub Zero Products
DistributionUS

Product

Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm or cool the blood that enters and exits a blood heat exchanger for adult, pediatric or infant patients.

Reason

Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.

Identifiers

code_infonone

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0629-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.