RecallRadar

FDA Device recall Z-0628-2016

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent

On 2016-01-20 the FDA classified a Class III recall of FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent by Becton Dickinson And Bd Biosciences, citing becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0628-2016
ClassificationClass III
Statusterminated
Initiated2015-12-14
Published2016-01-20
Last updated2026-09-13 11:54 UTC
CompanyBecton Dickinson And Bd Biosciences
DistributionUS

Product

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.

Reason

Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.

Identifiers

code_infoLot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0628-2016%22