FDA Device recall Z-0627-2016
Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagno…
- FDA Device
- Class II
- terminated
- Published 2016-01-20
On 2016-01-20 the FDA classified a Class II recall of Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagno… by Bayer Healthcare, citing the firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the d….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0627-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-30 |
| Published | 2016-01-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Bayer Healthcare |
| Distribution | US |
Product
Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.
Reason
The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail
Identifiers
| code_info | 50856405, 50857510, 50860994, 50866789, 50867921, 50869464, 50869465, 50871131, 50872965, 50878229, 50878230, 5087823…50856405, 50857510, 50860994, 50866789, 50867921, 50869464, 50869465, 50871131, 50872965, 50878229, 50878230, 50878231, 50878232, 50881328, 50881393, 50881394, 50881395, 50881506, 50881511, 50881512, 50882753, 50886316, 50886467, 50886468, 50886469, 50888285, 50888286, 50888287, 50888288, 50888373, 50890483, 50890484, 50890485, 50892357, 50892358, 50892359, 50892360, 60000321, 60000322, 60000323, 60000692, 60000789, 60000790, 60000833, 60000834, 60000835 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0627-2016%22