FDA Device recall Z-0626-2015
PKG, RING HANDLE, P/N 0250080236
- FDA Device
- Class II
- terminated
- Published 2014-12-24
On 2014-12-24 the FDA classified a Class II recall of PKG, RING HANDLE, P/N 0250080236 by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0626-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-17 |
| Published | 2014-12-24 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Endoscopy |
| Distribution | US |
Product
PKG, RING HANDLE, P/N 0250080236. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Identifiers
| code_info | 514021 614128 717447 814171 816251 816252 913442 913545 913581 913811 913812 913984 913985 914008 914371…514021 614128 717447 814171 816251 816252 913442 913545 913581 913811 913812 913984 913985 914008 914371 914372 914894 1016560 1016736 1016739 1016846 1016847 1113330 1113331 1113332 1114831 1116462 1116463 1216409 1314473 1314474 1314920 1314924 1314925 1315268 1315404 1316600 1316601 1316894 1316895 1316896 1410871 1410872 1410873 1410874 1411128 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.