FDA Device recall Z-0622-2016
UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilizatio…
- FDA Device
- Class II
- terminated
- Published 2016-01-20
On 2016-01-20 the FDA classified a Class II recall of UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilizatio… by Orthopediatrics, citing response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0622-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-10 |
| Published | 2016-01-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Orthopediatrics |
| Distribution | AL, AZ, CA, DE, FL, GA, HI, KY, LA, MI, MN, MO, MS, NC, NJ, NY, OH, OR, TN, TX |
Product
UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
Reason
Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw.
Identifiers
| code_info | Item number; lot; product description 00-1300-1735; 44003; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 35MM. 0…Item number; lot; product description 00-1300-1735; 44003; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 35MM. 00-1300-0735; 44601; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 35MM. 00-1300-0740; 44602; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 43926; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 44252; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0745; 43927; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 45MM. 00-1300-0750; 44619; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 50MM. 00-1300-0620; 44623; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 20MM. 00-1300-0640; 44617; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-0640; 44884; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-1745; 44005; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 45MM. 00-1300-0635; 44677; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 35MM. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0622-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.