FDA Device recall Z-0622-2014
Integra Large Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536…
- FDA Device
- Class II
- terminated
- Published 2014-01-15
On 2014-01-15 the FDA classified a Class II recall of Integra Large Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536… by Integra Lifesciences, citing integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0622-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-11-13 |
| Published | 2014-01-15 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Integra Lifesciences |
| Distribution | AL, CT, FL, MN, ND, NE, NH, NY, OK, OR, TX, WA, WI |
Product
Integra Large Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA NL 850-1132 Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.
Reason
Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to use but after the sterile package was open.
Identifiers
| code_info | 1094085, 1101479, 1101647, 1102403, 1102807, 1104174, 1110043, 1120817, 1130797, 1131331, 1131660 |
|---|