FDA Device recall Z-0621-2025
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro suscep…
- FDA Device
- Class II
- ongoing
- Published 2024-12-11
On 2024-12-11 the FDA classified a Class II recall of BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro suscep… by Becton Dickinson And, citing BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0621-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-11-06 |
| Published | 2024-12-11 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Becton Dickinson And |
| Distribution | US |
Product
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
Reason
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
Identifiers
| code_info | UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025). |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Becton Dickinson And
- BD Phoenix Panel NMIC-306
FDA Device · Class II · 2026-10-21 · Becton Dickinson And - BD Phoenix Panel NMIC-312
FDA Device · Class II · 2026-10-21 · Becton Dickinson And - ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code: 00690103210927 The non-invasive ForeSight oximeter cable is intended for us…
FDA Device · Class II · 2026-09-30 · Becton Dickinson And - BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in., BD¿Quincke Needle 22 G x 3.5 in
FDA Device · Class I · 2026-06-10 · Becton Dickinson And - BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in
FDA Device · Class I · 2026-06-10 · Becton Dickinson And
All Becton Dickinson And recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0621-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.