FDA Device recall Z-0620-2014

The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities ass…

On 2014-01-08 the FDA classified a Class II recall of The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities ass… by Ge Healthcare It, citing GE Healthcare is aware of a potential safety issue with the use of GE Centricity Laboratory Instrument Interface (IF) where free-text sent from IM is not being transcribed into fr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0620-2014
ClassificationClass II
Statusterminated
Initiated2013-03-12
Published2014-01-08
Last updated2026-09-13 11:53 UTC
CompanyGe Healthcare It
DistributionMN

Product

The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network. This software system will not be the software directly used to run, steer, or control any specific laboratory analyzer or equipment. It is intended for use in laboratory facilities, including a central laboratory or in a multiple laboratory environment servicing satellite laboratories, reference laboratories, clinics, etc. Centricity Laboratory System is not intended for direct patient contact. The Centricity Laboratory System product is designed to facilitate the general clinical, anatomic pathology and cytology laboratory workflow such as order entry, results entry, instrument interfacing, results reporting and patient record retrieval. It integrates with other Hospital Information Systems (HIS) through HL7 interface by order entering and also report processing. The Centricity Laboratory System is intended to interface with various lab results. Centricity Laboratory System is intended to be used by laboratory technologists, technicians, analyst and other trained/site authorized system users

Reason

GE Healthcare is aware of a potential safety issue with the use of GE Centricity Laboratory Instrument Interface (IF) where free-text sent from IM is not being transcribed into free-text in iNET. The instrument may upload results to IF/iNET when order/request is not yet in IF/iNET. In such a situation, test level supporting free-text sent from the instrument is discarded (not captured).

Identifiers

code_infoSoftware Versions: 3.3, 4.0, 4.1

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.