FDA Device recall Z-0609-2024
RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is designed to provide the determi…
- FDA Device
- Class II
- ongoing
- Published 2023-12-27
On 2023-12-27 the FDA classified a Class II recall of RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is designed to provide the determi… by Siemens Healthcare Diagnostics, citing RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0609-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-10-26 |
| Published | 2023-12-27 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491449
Reason
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
Identifiers
| code_info | UDI-DI: 00630414589770 All serial numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0609-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.