Ottobock Modular Polycentric EBS Knee Joint 3R60, 3R60=ST, 3R60=KD, 3R60=HD. Used for the prosthetic fitting of amputations of the lower. l…

On 2016-01-13 the FDA classified a Class II recall of Ottobock Modular Polycentric EBS Knee Joint 3R60, 3R60=ST, 3R60=KD, 3R60=HD. Used for the prosthetic fitting of amputations of the lower. l… by Otto Bock Healthcare, citing otto Bock HealthCare GmbH is recalling Modular Polycentric EBS Knee Joint that were serviced between 27 April 2015 and 17 November 2015 due to a service error where the wrong Loct….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0608-2016
ClassificationClass II
Statusterminated
Initiated2016-01-04
Published2016-01-13
Last updated2026-09-13 11:54 UTC
CompanyOtto Bock Healthcare
DistributionUS

Product

Ottobock Modular Polycentric EBS Knee Joint 3R60, 3R60=ST, 3R60=KD, 3R60=HD. Used for the prosthetic fitting of amputations of the lower. limb.

Reason

Otto Bock HealthCare GmbH is recalling Modular Polycentric EBS Knee Joint that were serviced between 27 April 2015 and 17 November 2015 due to a service error where the wrong Loctite was used during repair.

Identifiers

code_infoAll serviced between 27 April 2015 and 17 November 2015