FDA Device recall Z-0608-2013

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042

On 2013-01-02 the FDA classified a Class III recall of Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042 by Radiometer America, citing the barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0608-2013
ClassificationClass III
Statusterminated
Initiated2006-06-06
Published2013-01-02
Last updated2026-09-13 11:53 UTC
CompanyRadiometer America
DistributionKY, NC, NY

Product

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Reason

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Identifiers

code_infoLot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.