RecallRadar

FDA Device recall Z-0605-2015

PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171

On 2014-12-24 the FDA classified a Class II recall of PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171 by Stryker Endoscopy, citing the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethyl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0605-2015
ClassificationClass II
Statusterminated
Initiated2014-11-17
Published2014-12-24
Last updated2026-09-13 11:53 UTC
CompanyStryker Endoscopy
DistributionUS

Product

PKG, CANNULA 5.5MM AUTOMATIC VALVE STOP COCK W/O TROCAR, P/N 0250080171. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Reason

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Identifiers

code_info
35499 38198 38758 39650 39650N 40110 40250 43627 44955 44976 45866 46024 46320 46396 46710 46802 47210…35499 38198 38758 39650 39650N 40110 40250 43627 44955 44976 45866 46024 46320 46396 46710 46802 47210 47389 47513 47750 47970 48066 48398 48762 48965 49494 49498 49612 49834 50168 50430 50722 50855 50920 51536 52647 52919 53043 53136 53515 53795 53972 55872 58336 58821 59615 60227 60752 62482 64133 64678 64965 65640 67422 67737 68342 69272 71077 75257 75918 76080 76460 76831 76960 77711 78491 79425 82007 82916 83532 83926 85069 85467 85909 86149 86265 86430 86539 86869 87496 89088 89441

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0605-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.