FDA Device recall Z-0601-2019

Hemoglobin A1c Control, Part no. 220232 for use as quality control material to monitor the precision of laboratory testing procedures for H…

On 2018-12-19 the FDA classified a Class III recall of Hemoglobin A1c Control, Part no. 220232 for use as quality control material to monitor the precision of laboratory testing procedures for H… by Tosoh Bioscience, citing tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 pea….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0601-2019
ClassificationClass III
Statusterminated
Initiated2018-10-05
Published2018-12-19
Last updated2026-09-13 11:54 UTC
CompanyTosoh Bioscience
DistributionUS

Product

Hemoglobin A1c Control, Part no. 220232 for use as quality control material to monitor the precision of laboratory testing procedures for HbA1c quantitation.

Reason

Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.

Identifiers

code_infoLot 7055

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0601-2019%22