RecallRadar

FDA Device recall Z-0599-2025

Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plas…

On 2024-12-11 the FDA classified a Class II recall of Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plas… by Siemens Healthcare Diagnostics, citing potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) re….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0599-2025
ClassificationClass II
Statusongoing
Initiated2024-10-21
Published2024-12-11
Last updated2026-09-13 11:57 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635

Reason

Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.

Identifiers

code_infoUDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0599-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.