On 2018-12-19 the FDA classified a Class II recall of UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotra… by Neotract, citing upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0598-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-07-06 |
| Published | 2018-12-19 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Neotract |
| Distribution | AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV |
Product
UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
Reason
Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
Identifiers
| code_info | Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971,…Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971, P37985, P37988, P37995, P38001, P38013, P38020, P38124, P38132 |
|---|
Status history