RecallRadar

FDA Device recall Z-0594-2013

Zimmer Spine Ardis Inserter, Rx, non-sterile

On 2013-01-02 the FDA classified a Class I recall of Zimmer Spine Ardis Inserter, Rx, non-sterile by Zimmer, citing zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interb….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0594-2013
ClassificationClass I
Statusterminated
Initiated2012-11-12
Published2013-01-02
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

Zimmer Spine Ardis Inserter, Rx, non-sterile

Reason

Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up to 60 minutes could occur. Sharp edges of the broken implant may contact and damage adjacent structures, requiring surgical intervention. Observed health risks

Identifiers

code_info
Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71…Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71JE, 75LE Part, 07.00780.701, lot 76CN; part 07.00780.702, lot 76CP;part 07.00900.701, lot 76TR; 07.00900.702, 76TS; 650-0297-01, 70YT; 07.00780.701, 73EC; 07.00780.702, 73EC; 07.00900.701, 76TQ; 650-0239-01, 66PR; 650-0293-01, 70SG ; 650-0293-01, 70SG-R; 650-0179-01, 60CK; 650-0176-01, 60CL; 650-0365-01, 72LP; 07.00780.701, 76CQ; 07.00780.702, 76CR; 07.00780.701, 74EN; 07.00780.702, 74EP; 650-0351-01, 72JH; 650-0351-02, 72JJ; 650-0263-01, 68HL; 650-0208-01, 62FH; 07.00900.701, 76TN; 07.00780.701, 76QG; 07.00780.701, 76LC; 07.00780.702, 76LD;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0594-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.