RecallRadar

FDA Device recall Z-0588-2026

Quantum Intersegmental Table, Model/Catalog Number: AMQ400H

On 2025-12-03 the FDA classified a Class II recall of Quantum Intersegmental Table, Model/Catalog Number: AMQ400H by Pivotal Health Solutions, citing the heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0588-2026
ClassificationClass II
Statusongoing
Initiated2025-10-09
Published2025-12-03
Last updated2026-09-13 11:57 UTC
CompanyPivotal Health Solutions
DistributionUS

Product

Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table

Reason

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.

Identifiers

code_infoAll serial numbers.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0588-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.