FDA Device recall Z-0585-2015

ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical H…

On 2014-12-31 the FDA classified a Class I recall of ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical H… by Heart Sync, citing heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a co….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0585-2015
ClassificationClass I
Statusterminated
Initiated2014-11-11
Published2014-12-31
Last updated2026-09-13 11:53 UTC
CompanyHeart Sync
DistributionUS

Product

ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

Reason

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Identifiers

code_infoAll Lots. Catalog / Part Number T100LO-PHILIPS