RecallRadar

FDA Device recall Z-0578-2018

EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3

On 2018-02-21 the FDA classified a Class II recall of EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3 by Little Rapids, citing the secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient trans….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0578-2018
ClassificationClass II
Statusterminated
Initiated2017-12-27
Published2018-02-21
Last updated2026-09-13 11:54 UTC
CompanyLittle Rapids
DistributionCA, CT, FL, MO, MS, SD, TN, TX, WI

Product

EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3

Reason

The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.

Identifiers

upc10747036729303
code_infoLot codes: 17871, 17971, 18071, and 18971

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0578-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.