FDA Device recall Z-0576-2018
Contiplex Nerve Block Tray
- FDA Device
- Class II
- terminated
- Published 2018-02-21
On 2018-02-21 the FDA classified a Class II recall of Contiplex Nerve Block Tray by B Braun Medical, citing B. Braun Medical Inc. (BBMI) initiated this recall of various BBMI convenience kits containing the 22GA 1-1/2in SafetyGlide Needle due to the presence of loose polypropylene forei….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0576-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-06-21 |
| Published | 2018-02-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | B Braun Medical |
| Distribution | US |
Product
Contiplex Nerve Block Tray. Product Catalog Number: 570205. Local anesthesia of the skin prior to insertion of the regional anesthesia needle.
Reason
B. Braun Medical Inc. (BBMI) initiated this recall of various BBMI convenience kits containing the 22GA 1-1/2in SafetyGlide Needle due to the presence of loose polypropylene foreign matter was above release specification. This foreign matter has been identified as a product hub material which has been tested for biocompatibility per ISO 10993 during the product development process.
Identifiers
| code_info | 0061539146. 0061540984. |
|---|