FDA Device recall Z-0572-2016

CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance

On 2016-01-13 the FDA classified a Class II recall of CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance by Partners In Medicine, citing device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimatio….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0572-2016
ClassificationClass II
Statusterminated
Initiated2015-12-01
Published2016-01-13
Last updated2026-09-13 11:54 UTC
CompanyPartners In Medicine
DistributionUS

Product

CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic parameters time synchronized using Impedance Cardiography (ICG), Electrocardiography (ECG), oscillometric Blood Pressure (oscBP) and continuous Blood Pressure (contBP). Furthermore Sympathic Tone and Parasympathic Tone are calculated by power spectral analysis of ECG (heart rate variability) and continuous blood pressure (blood pressure variability) as well as the Baroreceptor Reflex Sensitivity (BRRS). For use by medically trained personnel. It has been designed for diagnoses aiding and must not be used for vital sign monitoring of critically ill patients under absence of additional, suitable self-monitoring devices or medical trained personnel. The device measures continuously the subject's hemodynamic parameters without reporting any diagnosis. It is the physician's responsibility to make proper judgment based on these parameters.

Reason

Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).

Identifiers

code_info
Serial numbers: 003040i-2014-161-014-GG-0000, 003040i-2015-101-003-GG-0000, 003040i-2014-161-001-GG-0000, 003040i-2…Serial numbers: 003040i-2014-161-014-GG-0000, 003040i-2015-101-003-GG-0000, 003040i-2014-161-001-GG-0000, 003040i-2014-161-002-GG-0000, 003040i-2014-161-003-GG-0000, 003040i-2014-161-004-GG-0000, 003040i-2014-161-005-GG-0000, 003040i-2014-161-006-GG-0000, 003040i-2014-161-007-GG-0000, 003040i-2014-161-008-GG-0000, 003040i-2014-161-009-GG-0000, 003040i-2014-161-010-GG-0000, 003040i-2014-161-011-GG-0000, 003040i-2014-161-012-GG-0000, 003040i-2014-161-013-GG-0000, 003040i-2014-161-015-GG-0000, 003040i-2014-161-016-GG-0000, 003040i-2014-161-017-GG-0000, 003040i-2014-161-018-GG-0000, 003040i-2014-161-019-GG-0000, 003040i-2014-161-020-GG-0000, 003040i-2014-161-021-GG-0000, 003040i-2014-161-022-GG-0000, 003040i-2014-161-023-GG-0000, 003040i-2014-161-024-GG-0000, 003040i-2014-161-025-GG-0000, 003040i-2015-101-001-GG-0000, and 003040i-2015-101-002-GG-0000

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0572-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.